Instability Risk Factors in Total Hip Arthroplasty for Femoral Neck Fracture
DOI:
https://doi.org/10.15417/issn.1852-7434.2022.87.5.1651Keywords:
Total hip replacement, instability, risk factors, hip fractureAbstract
Introduction: One of the most relevant complications of total hip arthroplasty (THA) in the treatment of fractures is a dislocation rate of 10%, which is 5 times higher than that of THA for osteoarthritis. Our objective was to determine the dislocation rate in patients treated with THA due to femoral neck fracture and to evaluate the cause of dislocation.
Materials and Methods: We carried out a retrospective study on 110 THAs in patients with femoral neck fractures (the average age was 69 years, and 71% were women). The type of fixation of the prosthesis, the size of the head, the offset, and the surgical approach were evaluated. The angle of acetabular inclination and anteversion and the comorbidities were measured.
Results: 72% of surgeries were performed through the posterolateral approach. The anterolateral approach is associated with a 7° more vertical cup (p=0.001). 65.4% of the heads were 32 mm. 15% of the small heads (22 and 28mm) (3/20) and only 1% of the large heads (1/90) dislocated (p=0.0027). No 36-mm heads were dislocated. Depression, Parkinson’s, and Alzheimer’s disease were the most frequent neurological comorbidities.
Conclusions: Small-diameter heads, in combination with poor positioning of the prosthesis and neurocognitive diseases, are associated with greater instability. Using prosthetic heads with a diameter of 36 mm and a correct orientation of the components is sufficient to ensure stability.
Downloads
Downloads
Published
License
Manuscript acceptance by the Journal implies the simultaneous non-submission to any other journal or publishing house. The RAAOT is under the Licencia Creative Commnos Atribución-NoComercial-Compartir Obras Derivadas Igual 4.0 Internacional (CC-BY-NC.SA 4.0) (http://creativecommons.org/licences/by-nc-sa/4.0/deed.es). Articles can be shared, copied, distributed, modified, altered, transformed into a derivative work, executed and publicly communicated, provided a) the authors and the original publication (Journal, Publisher and URL) are mentioned, b) they are not used for commercial purposes, c) the same terms of the license are maintained.
In the event that the manuscript is approved for its next publication, the authors retain the copyright and will assign to the journal the rights of publication, edition, reproduction, distribution, exhibition and communication at a national and international level in the different databases. data, repositories and portals.
It is hereby stated that the mentioned manuscript has not been published and that it is not being printed in any other national or foreign journal.
The authors hereby accept the necessary modifications, suggested by the reviewers, in order to adapt the manuscript to the style and publication rules of this Journal.